search Search

Onboarding

Starting mylife Loop!The mylife Loop system comprises three integrated components from three separate manufacturers:

  • The YpsoPump insulin pump, used in conjunction with a compatible infusion set

  • The CamAPS FX application

  • A compatible continuous glucose monitoring (CGM) system

Comprehensive training on each individual component is required prior to system initiation. Users must demonstrate appropriate understanding of device operation, infusion set management, application configuration, and CGM utilisation.

In addition, users should be clearly informed about manufacturer-specific support pathways and how to access technical and clinical assistance where required.

Please use the structure and supporting materials outlined below to guide and document a safe and effective mylife Loop start

1. What is required to complete a mylife Loop start

  • Pump Training Checklist must be completed during the pump start and recorded in a dedicated storage drive for future reference

  • YpsoPump with Instructions for Use (IFU)

  • Compatible phone with CamAPS FX app downloaded with in-app training completed (check the phone is compatible with chosen sensor)Dexcom G6FreeStyle Libre 3/Libre 3 Plus

Compatible sensor:

  • Infusion set and insulin cartridge/pump cart

  • Basal settings, bolus calculator settings, weight, total daily dose and personal glucose target settings (must be agreed with clinical team before the pump start). Our set up guide with help with these.

  • Any pre-pump start training tasks which have been advised by clinical team or pump trainer

attachment
A handy resource on the mylife Loop specifications and for general information:
download mylife Loop powered by CamAPS FX - Specifications (1.5 MiB)

2. Instruct user how to use YpsoPump

The YpsoPump will deliver insulin to the body via the insulin in the infusion set. It is vital that users have a good understanding of using the YpsoPump insulin pump with and without the CamAPS FX app.

  • The user guide (IFU) which comes in the YpsoPump box contains information on all the functions of the insulin pump. Using the training checklist please ensure all topics are trained successfully. The IFU is also a great resource for users to refer to at home if needed.

  • The IFU can also be access on our website named “user guide”

  • There is a quick start guide for the YpsoPump which is a great resource when teaching the functions of the YpsoPump and compliments the checklist.

  • The user must have agreed with the clinical team on basal rate settings for the YpsoPump which will be input into profile A and B in the basal setting. The set up guide can be a helpful tool for this.

  • YpsoPump uses a standard AAA battery. Ensure the user knows the low battery alerts and alarms, how to change the battery and the related risks, and also to always carry a spare battery.

NOTE:

  • The user must understand the main menu of the pump and where to find functions and settings. A map of the menu is available in Chapter 10 of the IFU and in the quick start guide.

  • The user must ensure that there are no issues with insulin delivery from the YpsoPump, all pump alerts must be acknowledged, regular infusion set changes 2 to 3 days and ensuring there is enough insulin in the YpsoPump at all times.

  • Ensure the user is familiar with all of the icons and alarms, a legend of these can be found in Chapter 10 of the IFU.

  • A virtual YpsoPump app is available on the Android and iOS app store called Explorer app. A patient can practice using all the functions of the YpsoPump using a 3D simulator as well as access guided tours and a glossary of pump icons and alarms.

3. Guide user on infusion sets and completing a set change

A user must be very confident in conducting an infusion set-change and inserting their chosen cannula before using the YpsoPump. This is one of the most important training aspects, as if the infusion set is not functioning, the insulin delivery will be compromised. There must be a low threshold for changing the infusion set in the event of suspected delivery issues.The Ypsopump has 3 infusion set options

  • Inset infusion (Teflon cannula, available in tube lengths from 46cm to 110cm and cannula length 6mm and 9mm)

  • Orbit micro (steel cannula, available in tube lengths from 46cm to 110cm and cannula length 6mm and 9mm) can be used with the mylife orbit inserter link

  • Orbit soft (Teflon cannula, available in tube lengths from 46cm to 110cm and cannula length 6mm and 9mm)

When priming, it is vital for the user to know the length of the tube and the cannula to know the priming volume required. The information on the tube and cannula length is on the side of the box and on the infusion set packaging. This information can be found in the IFU in the box of infusion sets or on the website.

Reminder: Encourage the user to always check which tube and cannula length and know what is normal for them. Advise them to contact for support if they are sent an alternative cannula to what they usually use.

attachment
Helpful resources when teaching infusion set changes
download mylife YpsoPump Reservoir - Flexibility with your insulin of choice (2.4 MiB)

Refilling process and completing a set change

The insulin reservoir can be refilled once. Please ensure the user does not refill the reservoir unless they feel confident doing so. Ensure the risks of refilling the reservoir more than once are advised. Encourage the user to follow this video for the refilling process.

Using YpsoPump without CamAPS FX app

Some YpsoPump functions which are particularly important for users to know in the event that the CamAPS FX app is not functioning or using the YpsoPump alone are:

  • Bolusing from the YpsoPump without a bolus calculator (manually calculating corrections and meal boluses) can be found in Chapter 4 of the IFU. This is particularly important in the event that the phone is lost, and the app cannot be accessed. Note that even in open loop a bolus can be given via the CamAPS FX app.

  • How to use temporary basal rates can be found in Chapter 3 of the IFU

  • Changing basal rates is also in Chapter 3 of the IFU

4. Setting up and using CamAPS FX app

Once the user is confident on how to use the insulin pump and familiar with the functions and infusion set changes, turn focus to education on HCL and pair the YpsoPump insulin pump to the CamAPS FX app.

attachment
This is a helpful resource to help guide teaching the user on how to use the app

In-app training must be completed for the app to function; the user is prompted to go through each animation in a sequence before being able to use the app. It is advised to encourage the user to do the inapp training in their own time before the AID start.

The user will also be prompted to add the bolus calculator settings, weight and TDD during set up. The PGT is defaulted to 5.8 mmol/L (105 mg/dL) if not manually adjusted.

The CamAPS FX user manual can be found in the help section of the app or on the CamDiab website.

Training factsheets on the following CamAPS FX functions are available on the app: bolus calculator, using “Boost”, using “Ease-off”, Calibration, using “Add meal”, app alerts, Data share options, Accessing Statistics, Block, Changing settings.

Access fact sheets on CamAPS FX app via: main menu – help – in app training – Download training factsheet.

NOTE:

  • The in-app training is usually done prior to the YpsoPump start by the user so they have time to focus on all the necessary animations and information. The training may also be done in clinic with the user but it must be completed before the AID system can be used. A user can go back at any time and watch the in-app training animations again. The in-app training only covers the functions in the app not the YpsoPump.

  • Using "Boost" will not prevent ketone development if there is an infusion set issue. Basics insulin pump upkeep still apply.

  • The main setting which will directly effect the automated insulin delivery is the PGT. The bolus calculator settings only effect manual bolus and corrections.

  • Hybrid Closed loop requires accurate CGM data, if the CGM is reading inaccurately for a period of time, the glucose can be calibrated on the app to correct this. Ensure the user carries a blood glucose meter at all times in the event of a CGM failure or suspected inaccurate readings.

5. Setting up follow and Share options

mylife Loop has many data share options for caregivers and clinicians to access the users glucose and insulin data as well as notifications and alarms. A factsheet on sharing diabetes data can be accessed on the app under the help section – in-app training – download factsheets. Ensure a caregiver has the correct data sharing and alert notifications enabled at the YpsoPump start.

  • Companion app – The CamAPS FX app can be downloaded by caregivers on their own phones to act as the companion app which mirrors the users app. No insulin adjustments can be made with the companion app but all data which can be accessed on the users app can also be seen on the companion app. Find out more about the companion app in the in-app training videos access in the help section of the app.

  • Glooko and mylife Cloud – these programs are usually access by the clinical team, please ensure the accounts are set up and shared with the clinical team at the YpsoPump start.

  • SMS – all alerts from the CamAPS FX app can be sent via SMS to up to 5 followers. The user must have a working SIM card in the phone for this function to work.

6. Vital supplementary information for specific population groups

Pregnancy – Pregnant women may have an increased risk of complications related to Type 1 Diabetes and must react promptly to out of range glucose levels. It is imperative that pregnant women are aware of these factors, and have the correct tools to mitigate these risks, for example appropriate sick day rules, managing high glucose and ketones, emergency situations, as well as support from friends or family members.

attachment
Ensure that pregnant women receive copies of the specific pregnancy information
download Managing pregnancy on mylife Loop (10.5 MiB)

Also information available on the CamAPS FX app in; In app training video and app Factsheets on the CamAPS FX app under help in the main menu.

Children – Due to their unpredictable routines and activity levels, children’s glucose levels may fluctuate more than an adults. Use the information found here to highlight some of the functions in the mylife Loop that can help support families in their daily lives.

attachment
Ensure that caregivers receives copies of the specific information on children
download Optimising the mylife Loop AID system, powered by CamAPS FX (10.5 MiB)

Adolescence – Some functions on the CamAPS FX app may assist with the challenges in this age group.

attachment
Specific resources for adolescents
download Clinical Insights - Adolescents (18.9 MiB)

Check your Understanding

Use the list below to encourage the user to use the additional resources to become confident in using important functions in the AID system and know where to find the information.

YpsoPump functions

Function

Additional resources if user requires more information

Know how to power on/off the YpsoPump and screen unlock

Explorer app or IFU Chapter 2 (in YpsoPump box and website)

How to navigate the YpsoPump and

understand of menu icons

YpsoPump e-learning or IFU Chapter 2

YpsoPump Handling videos

How to program basal rates

IFU Chapter 3

Where on the YpsoPump to switch basal profiles

IFU Chapter 3

Use of temporary basal rate

IFU Chapter 3

How to deliver a bolus via the YpsoPump

IFU Chapter 4

Understanding bolus types

IFU Chapter 4

Be able to identify the YpsoPump alerts and alarm symbols

IFU Chapter 8

Infusion set management

How to change Cartridge

Inset Orbit2Micro Orbit2Soft

IFU Chapter 5

Infusion set change

Infusion set change

Handling videos

IFU Chapter 5

Identification and use of different cannula types

Infusion sets

YpsoPump troubleshooting

IFU Chapter 8

Risks associated with insulin pump therapy and other activities

Hypoglycaemia

Hypo Factsheet

Risk of DKA using an insulin pump

DKA-in-pregnancy

DKA-in-children

DKA-in-adults

Using an insulin pump during illness

Managing illness with mylife Loop

Using an insuling pump when travelling

Travelling with mylife Loop

Using an insulin pump during Excercise

Utilising mylife Loop powered by CamAPS FX when exercising with type 1 diabetes

In app training video

Meal Management

Managing meals with mylife Loop

During surgery

Managing mylife Loop during surgery

CamAPS FX app

Understanding core functions of the CamAPS FX app; Bolusing, "Boost", "Ease off", "Add Meal", Simplified meal announcement

In app training and app Factsheets

How to change settings on the app?

In app training and app Factsheets

Sharing data from the app; Companion app, Glooko, SMS alerts, CamAPS FX PDF report

app Factsheets

Calibrating the sensor

Optimising the mylife Loop

Optimising mylife Loop

Back-up and contingency planning

Availability of back-up insulin pens

Ask your diabetes team for copies of your sick day rules and ensure you have an up-to-date prescription/supply of required items. Have any clinical helpline phone numbers saved and accessible to all family members.

Availability of glucose meter

Availability of ketone meter

Emergency management plan reviewed

Emergency contact numbers confirmed

mylife Customer care for mylife Loop related issues

Customer Care & Contact